Hoechst or DAPI for nuclear staining? When is it best to use Hoechst vs. DAPI for nuclear staining? They seem to be very similar on paper. Are there situations where one is clearly preferable?

The Hoechst 33342 dye is similar to DAPI in that both are UV-excited, minor groove-binding, and emit signals proportional to total DNA content. Both are maximally excited around 355 nm and emit around 460 nm. A UV light source is required, which may harm the cell. However, this is only a risk with Hoechst, as DAPI requires the cells to be fixed and/or permeabilized, which is incompatible with life. DAPI is more stable, but Hoechst is brighter. Both are subject to photo-bleaching after long exposure.

There is another newer DNA dye called DRAQ5 that makes up for all of the issues I mentioned earlier - it is bright, quite photo-stable, can be used in live cell imaging, and its excitation and emission wavelengths are in the far red end of the spectrum, meaning that no UV is needed. I found a good comparison chart at < if you're interested. I have no connection to them, by the way, I've just used DRAQ5 a lot in the past.